Drager Narkomed MRI

One vaporiser for vaporising 19.1 monitor style and volume, pressure and oxygen in CORM first generation drives. The second generation units have a Drager GS type screen volume control, pressure and oxygen. Absorber assembly has the feeling Narkomed Known but MRI compatible version. The units are small and Have Been tested near the strongest magnets.

  • Fan: Pneumatic double circuit ascending fan + fan AV2 style without compensation tidal volume. Volume Control modes (VCV), Manual / spontaneous. Mechanical bag switch / ventilation.
  • Checklist machine: Manual check list of the FDA.
  • Flowmeters: With traditional needle valve flow display glass.
  • Breathing circuit: Head traditional absorbent. The breathing circuit is of greater volume (absorb two vessels). You can use loose granules or preloaded containers. It can be used formally no re-breathing circuits.
  • Power failure: Most functions except patient monitoring, are retained During a power failure as it is a basic pneumatic-mechanical machine.

Description

general

  • Dimensions:  (W x H x P) 31.5 “x 53” x 27 “(80 x 135.3 x 68.6 cm)
  • Weight:  230 lbs (104.3 kg)
  • Weight battery charger:  3 lbs (1.36 kg)

Electricity

  • 120 Volt Power Supply 100-120 VAC @ 50/60 Hz
  • 220/240 Volt Power Supply 200-240 VAC @ 50/50 Hz

Battery

  • Reserve Power Time “3 hours
  • Charging time “16 hours

Gas delivery system

  • 3 Capacity number of gas gases (O2, N2O, Air)
  • Flow 55 L oxygen discharge / min
  • Low oxygen supply
  • Alarm pressure 34 – 40 psi

Vaporizer

  • Vapor 2000
  • 2 positions
  • Hardmount

Fan

  • AV 2+
  • Frequency 1-99 (in increments of 1 BPM)
  • I: E ratio 4: 1 to 1: 4.5
  • Inspiratory Flow 10 -100 L / min
  • Tidal volume 20 mL -1500
  • Adjusting the pressure limit 15 cmH2O -120

Oxygen Monitoring

  • Display range 10 -100%
  • 1 vol% O2 Resolution

Monitoring airway pressure

  • Display range from -10 to 125 cmH2O
  • Resolution 1 cmH2O

Monitoring tidal volume

  • Display range .01 – 2.0 L
  • Resolution 0.01 mL

Utility requirements

  • The system works with a 120 VAC at 50/60 Hz.
  • regulator assemblies and external cylinder heads for O2 and N2O.
  • O2, N2O and air hoses to connect the sources of pipe connections DISS machine.

Tests MRI scanners up to 3.0 Tesla (shielded or unshielded)

  • With the unit performing tests MRI.