  • Proven versatility ventilation
  • Electronic Gas-Flow Measurement
  • Breathing system compact and convenient
  • Intelligent Cable Management

Description

Dräger Fabius GS Specifications

Weight:

  • 224 lbs (101.6 kg) (base unit without vaporizers or cylinders).

Dimensions:

  • (W) x 89.5 cm (H) x 130 cm (D) 82 cm (35.2 x 51.2 x 32.3 inches)

Power supply:

  • 100 – 240 VAC, 50/60 Hz, 2.3 max.
  • Battery (supports fan and monitor)> 45 min.

Fan:

  • E-vent Electrically controlled, electrically actuated.

Standard operating modes:

    • Manual / Spontaneous.
    • Volume Control (IPPV).
options:
    • Pressure control. (PCV)
    • Pressure Support. ()
    • Volume controlled ventilation synchronized With PS. (SIMV / PS)

Breathing frequency:

  • 4 to 60 lpm.

Max. Minute volume (MV):

  • 25 l / min.

PEEP:

  • 0-20 cmH2O.

Inspiratory / Forfeiture:

  • (Ti: Te) 4: 1 to 1: 4.

Pressure limit (Pmax):

  • 15-70 cmH2O.

Tidal volume (Vt):

  • 20-1400 mL in volume control.
  • 20-1100 mL SIMV / PS.

Inspiratory Pause (Tip Ti):

  • 0-50.

SIMV inspiratory time (Tinsp):

  • 0.3 to 4.0 sec.

Inspiratory pressure (Pinsp):

  • CmH2O PEEP + 5-65.

inspiratory flow (InspFlow):

  • 10-75 L / min in volume and pressure control.
  • 10-85 L / min in pressure support.

Pressure support level (Δ PPS) PEEP:

  • + 3-20 cmH _ {2} O 3 Frequency Min -. 20 bpm and “OFF”.

Apnea-Ventilation:

  • (Min Freq ..) Trigger 2-15 L / min.

Security features:

  • The Oxygen Ratio Controller Sensitive (S-ORC) Guarantees a minimum O2 concentration of 23% in a mixture of O2 / N2O. N2O if the fresh gas valve is closed O2 or O2 if the flow rate is less than 0.2 L / min.

Security features:

  • The Oxygen Ratio Controller Sensitive (S-ORC) Guarantees a minimum O2 concentration of 23% in a mixture of O2 / N2O. N2O if the fresh gas valve is closed O2 or O2 if the flow rate is less than 0.2 L / min.
  • audible and visual indication (flashing red LED) if the O2 pressure drops below 20 psi (1.38 bar) ± 0.9 bar (4 psi). In case of failure of electricity and battery, manual ventilation, gas delivery and delivery gent are possible. The relief valve opens positive pressure to 75 ± 5 cmH2O. The relief valve opens negative pressure to – 7.5 to – 9 cmH2O.

Range of indicators fresh gas flow:

  • 0.00 to 12.0 L / min.

Total fresh gas flow meter:

  • 0 to 10 l / min, calibrated with a mixture of 50% O2 and 50% mixture of N20.

O2 to 87 psi (6 bar):

  • Max. 75 L / min. 41 psi (2.8 bar): min 25 l / min.

Vaporizer assembly:

  • Dräger or SelectatecR.

Supervision:

  • Continued monitoring of the inspiratory O2 concentration (may be disconnected for service), respiratory rate, tidal volume, minute volume, medium pressure or plateau, peak airway pressure and PEEP. In Addition, all fresh gas flow information is displayed as virtual flow tubes.

Serial interface:

  • 1 x RS 232 (standard); 1 x RS 232 (optional)