Description
Dimensions
• Height x Width x Depth: 14.6 x 42.5 x 25.4 cm (5.75 x 16.75 x 10 inches)
• Weight: 3.6 kg (8 pounds)
energy requirements
• Rated line voltage: 100-230 VAC
• 50/60 Hz line frequency (47-63 Hz Operates on)
• Power consumption: ≤ 30 VA
• Monitor DC Input 12 Vdc at 2.5A
Operating Specifications
• Pulsed Doppler processing Technical automatic correlation
• Transducer type 9-crystal
• 2 kHz pulse repetition (all modes)
• Frequency Pulse duration 92 microseconds
• 1151 MHz frequency transmitter
• Isata Temporary Space-Media <5 mW / cm2
• Medium intensity
• Beam 20 dB focal area 16.6 cm2, range = 7 cm
• instantaneous peak 1.8 mW / cm2
• Intensity
• Heart rate range 50-210 BPM count
• Leakage current Meets IEC standard 60 1.1 harmonized national and / or IEC 601.1.1; Meets EN 60601-1-2: 2002
fecG mode
• Technique: Peak Detect, cardiotachometer beat-to-beat
• Cardiac interval range: 30-240 BPM
• Heart rate resolution: 1 BPM
• Artifact Removal: selectable service ± 25 BPM artifact rejection
• Accounting signal range input: 15μV to 2 mV peak to peak
• Tolerance compensation voltage (differential): ± 300 mVdc maximum
• high common mode voltage: 20V peak to peak
• Common Mode Rejection: Balanced:> 120 dB at mains frequency, with patient cable 5kΩRA or LA:> 110 dB at
mains frequency
• Input Impedance: Differential:> 10MΩ; Common mode:> 20MΩ
• Mains frequency rejection:> 40 dB
• Leakage current: Meets the standard IEC 601.1 harmonized national and / or IEC 601.1.1
• Isolation network to Patient:> 5656 Vdc
Mode ultrasound
• Technique: Pulsed Doppler With autocorrelation process
• Transfer type transducer 9-crystal
• Pulse repetition frequency: 2 kHz (all modes)
• Pulse duration: 92 ms
• Transmitter Frequency: 1151 MHz
• Average space-temporal average intensity: Isata <5 mW / cm²
• 20 dB beam focal area: 16.6 cm², range = 7 cm
• maximum instantaneous intensity: 1.8 mW / cm²
• Heart rate: 50-210 BPM
• Leakage current: Meets the standard IEC 601.1 harmonized national and / or IEC 601.1.1