Hospira Plum 360

With today’s hectic pace, tight budgets and Patient Safety Concerns, needs an infusion pump That serves you and your Patients Both, helping to maximize valuable resources and Improve Their clinical practice. You will find That the system 360 Plum infusion is unlike any other, with a unique combination of features him to Provide That Enables optimal health care – today and tomorrow.

Description

  • Air Matters – Plum 360 the exclusive use cassette PlumSet ™ offering air management does not require disconnecting That the patient
  • Remove the air faster – Automatically use the pump to remove air through the secondary port administration cassette without disconnecting the patient line
  • Reduces interruptions of therapy – capture 1 mL of air in the air trap of the cassette – a unique feature of the industry – before activating the alarm
  • Reduce the risk of contamination – help eliminate potential contaminants maintaining the patient line connected
  • Secondary distribution is paramount – The infusion systems pen Hospira are unique in how they connect the secondary line directly to the cassette PlumSet, so that the pump can differentiate between the lines of primary and secondary infusion and delivering them as The intention is to
  • Optimize security and compliance at the point of care – to program your infusion system 360 Plum with Hospira MedNet ™ software automatically starts in the drug library
  • EMR Ready, Today and Tomorrow – when you and your partner Electronic Medical Record (EMR) are ready for integration, Plum infusion system 360 with Hospira MedNet software security is also ready. Either today or at any time thereafter, the integration solution Clinic IV Hospira Plum 360 will connect your system to order medication, patient information and EMR

Hospira Plum 360 Specifications

Dimensions

  • Dimensions approx. 8 “H x 8” W. X 6 “D (0,20 m x 0,20 m H L x 0.15 m), excluding the clamp and storage pole cord
  • Approximate weight. 10 lbs (4.5 kg) with battery

Power

  • Battery life Approx. 3 hours to 125 mL / h in a single line, or at least 250 mL delivery if> 126 mL / h
  • Power requirements 120 V, 50 to 60 Hz, 50 VA
  • Recharging time Approx. 8 hours with the device running at 125 mL / h in a line

Delivery Rates

  • Lines A and B: 0.1 to 99.9 mL / h (in 0.1 mL), 100-999 mL / h (in 1 mL)
  • Concurrent Delivery: 0.5 mL / hr minimum for each line
  • PlumSet: 500 ml / h cumulative (A + B) maximum
  • KVO: 1.0 mL / hr or last delivery rate in the associated line, whichever is less

Precision delivery rate

  • ± 5% (1 to 999 ml / h), ± 10% (0.1 to 0.9 ml / h)

Delivery units

  • Mg / m2 / min, mg / m2 / h, mg / kg, mg / kg / h, m / g / min, mcg / min, mcg / kg / h, mg / mg / h, ng / kg / min, g / h, mEq / h, Mun / min, units / min, units / hr, units / kg / h

general

  • Plum pump line 360 ​​™
  • General Infusion Pump type
  • Volume range of 0.1 to 99.9 mL (0.1 mL increments), 100-9999 mL (in 1 mL)
  • Alarm system, Air occlusion, low battery, door open During pumping Rape full locking VTBI, valve test / cassette, nurse call, No Action, incomplete Programming
  • Occlusion pressure alarm limit distal pressure (no alarm): 1-15 psig (default 6 psig)